Table 5 Spearman’s rank-order correlation between ROTEM and Quantra

From: Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study

Comparison

r (95% CI) sample 1 (n = 37)

r (95% CI) sample 2 (n = 35)

r diff sample 1 to sample 2 (95% CI)

r diff p-value*

EXTEM A10 (hPa) vs Quantra CS (hPa)

0.93 (0.86 to 0.96)

0.94 (0.88 to 097)

− 0.01 (− 0.08 to 0.05)

1.00

FIBTEM A10 (hPa) vs Quantra FCS (hPa)

0.92 (0.85 to 0.96)

0.96 (0.92 to 0.98)

− 0.04 (− 0.11 to 0.01)

0.60

INTEM CT (sec) vs Quantra CT (sec)

0.71 (0.51 to 0.84)

0.75 (0.55 to 0.86)

−0.03 (− 0.28 to 0.20)

1.00

HEPTEM CT (sec) vs Quantra CTH (sec)

0.59 (0.33 to 0.77)

0.77 (0.59 to 0.88)

−0.18 (− 0.47 to 0.08)

0.68

(EXTEM A10 – FIBTEM A10) (hPa) vs Quantra PCS (hPa)

0.15 (−0.18 to 0.45)

0.80 (0.64 to 0.90)

−0.65 (− 0.99 to − 0.33)

0.0002

Platelet count (×1000/μL) vs (EXTEM A10 – FIBTEM A10) (hPa)

0.70 (0.48 to 0.83)

0.69 (0.46 to 0.83)

0.01 (−0.23 to 0.25)

1.00

Platelet count (×1000/μL) vs Quantra PCS (hPa)

0.37 (0.05 to 0.62)

0.69 (0.46 to 0.83)

−0.32 (− 0.65 to − 0.01)

0.28

  1. 95% CI, 95% confidence interval, r diff; difference between regression coefficient of sample 1 and 2
  2. For ROTEM parameters EXTEM and FIBTEM, A10 refers to amplitude at 10 min
  3. Quantra parameters: CS, clot stiffness; FCS, fibrinogen contribution to clot stiffness; PCS, platelet contribution to clot stiffness; CT, clot time; CTH, clot time with heparinase
  4. * P-values for the comparison of correlations between samples were calculated according to Zou and Silver et al. [16] and adjusted for multiple testing by the Holm-Bonferroni method [17]